Preparing for the Surveys

  1. Print “Lab List Updated mm-dd-yyyy.xlsx” located in ORS Files/Campus Survey

    • Ensure that the list of labs is updated including
      • comprehensive list of unsealed source labs
      • footnotes regarding status of labs is current
  2. Generate preliminary inspection report from the EHSA Database

    1. In the EHSA database select the “Inspections” tab
    2. In the “Inspections” box, click “RAM Inspections” under “Reports”
    3. Select “Preliminary Inspection – Generic”
    4. For the start and stop date, use the date you are running the report
    5. Enter one or more PI’s using the information from the “Lab List Updated mm-dd-yyyy.xlsx”
      1. type the last name to quickly find a PI.
      2. Use the up/down arrows on the screen or keyboard to navigate the list
      3. hit enter to select the highlighted PI or click on the PI
      4. repeat steps i-iii for all PI’s
      5. Print the report for review prior to visiting the lab
        1. Use this report specifically to
          1. review any unsatisfactory items from the previous inspection to ensure they have been corrected
          2. see if any of the personnel are overdue for Refresher Radioactive Material Safety
          3. check that the total number of source vials in storage does not exceed the number shown in the “Specific Nuclide Inventory” section
          4. compare listed survey meters with those physically present in lab; ensure within 1 year calibration period; turn on an confirm operable
  3. Print Survey Maps in ORS Files/Campus Survey

    1. Survey maps are organized by the machine that the smears are run on (LSC, LB5100, Gamma Counter); RS-117 (LB 5100 Rooms).xlsx, RS-117G (Gamma Rooms).xls, RS-118 (LSC Rooms).xlsx
      1. Instruments used are based on sources that the PI is authorized for which can be verified against the “Preliminary Inspection - Generic” report
    2. Yellow tabs indicate that the lab is on the lab list printed in Step 1.
    3. Print all lab maps for labs that are on the lab list from section 1.
  4. Filter Paper Smear Preparation

    1. Place numbered smears in envelopes for labs requiring a contamination survey and attach to Survey Map with a paper clip.
    2. Use the same number of smears as the previous map unless previous map was a de-activation survey, which generally has more.  If unsure, ask an ORS staff member.
  5. Adding the Inspections to EHSA in the EHSA database, schedule an inspection for each lab that it is required (inspections in the same building for the same PI should be combined in one inspection)

    1. In the EHSA database select the “Inspections” tab
    2. In the “Inspections” box, click the “Schedule Inspections” link
    3. In the “Inspection Details” box
      1. In the “Inspection Type” pulldown enter or select RAM
      2. Select “View Labs by PI”
      3. Type last name of PI
      4. In “Inspector” Dropdown enter RSO (Steve Grimm). (note, this will be reassigned at the time of the inspection)
      5. In “Permit Type” select RAM
      6. Click the “Search” button
      7. For PI’s with multiple labs, if the labs are in the same building, select the checkbox in the “Combine” column on the appropriate rows. (note do not combine separate buildings, because the inspection may be done by a different inspector on a different day)
      8. For PI’s with a single lab, select the checkbox in the “Single” column
      9. Click the “Schedule/Create” button
  6. Supply Cart Preparation

    1. An appropriately sized lab coat for each inspector
    2. Boxes of nitrile glove appropriately sized for each inspector
    3. Large Waste Bag
    4. Small Waste Bags
    5. GM and, if needed, ion chamber.
    6. Tablet computer (or Inspection Checklist)
    7. Extra Smears
    8. Extra Smear Envelopes
    9. Safety Glasses
    10. Shoe Coverings
    11. Cotton “Q-tips”
    12. Postings
      1. RAM stickers
      2. No Eating or Drinking stickers
      3. DNR Emergency Contact
    13. Manila Folder
      1. Campus Survey Smear Key
      2. Small Radiation Stickers
      3. Replacement Source Number Stickers
      4. Notice to Employees
    14. Scotch tape and/or double-sided tape for postings

Performing the Surveys and Inspections

  1. Conducting a Survey

    1. If two people, one person can do the inspection report, while the other collects the smears.
    2. Inspection report
      1. In EHSA database, click the “Inspections” menu icon, and then in the “Inspections” box, select “Inspections by Inspector”.
      2. If you are returning to a previously started inspection, Leave the “Inspector:” drop-down as your name
      3. If you are starting a new survey, select RSO (Steve Grimm) as the “Inspector:”
      4. The “Category” dropdown should have RAM selected.
      5. The “Status” dropdown should have Pending selected.
      6. Click the “Search” button
      7. Find the Building that has the inspections that you are conducting. Click the “View Inspections” button.  Some buildings, have multiple inspections so you may have to click to verify that you are on the right inspection
      8. Double-click on the line item with the correct rooms
      9. On the first page:
        1. Update “Inspection Date” to the current date of the inspection.  Enter the same date in the "Date Completed" field.
        2. Change “Lead Inspector” to your name.  This is the person filling out the inspection checklist
        3. Change “Inspector” to your partner’s name, if necessary.  This is the person doing the contamination survey.
        4. If a lab member is present and participated in the inspection with you by answering questions, etc., enter their name in the
          “Accompanied By” field.
        5. Click “Save and View Checklist”
      10. All RAM labs must be inspected for the “Basic Requirements”. To start click the “Basic Requirements” disclosure button to view the items.
        1. Each line item must have a response in the “Answer” column, which is the 6th column.
        2. For the postings in a neighborhood lab, they should be present on any door that leads into the neighborhood area
        3. If the Lab is enclosed and locked, then the “Neighborhood” lab questions should be “N/A”
        4. If the lab is a neighborhood lab, then the “Enclosed” lab question should be N/A
        5. If the lab is inactive and no unsealed sources present, then the “Contamination Survey” question is N/A
        6. The “High Activity Sources Only” question should be N/A.  This question is only answered when an inspection is being conducted in conjunction with Work Order 337, Survey/Inspect High Rad Sources.
        7. For items found unsatisfactory, record what was seen as unsatisfactory in the field “Observations” and record the recommended corrective action to fix it in the field “Recommendation”.
        8. If any item is Unsatisfactory, the inspection is for a PI with multiple labs, and the unsatisfactory applies to only some of the labs:
          1. Select the row for the unsatisfactory item
          2. Duplicate the item for each of the unsatisfactory labs
          3. Answer the same question for each lab as Unsatisfactory
          4. For the unsatisfactory lab(s) select the applicable room in the column “Lab/Building Name”
          5. If any of the labs was satisfactory for the given item, be sure to mark it as Satisfactory and leave the column "Lab/Building Name" blank.
      11. Answer the questions in the “Active Use of RAM” section as applicable.
      12. If work in the lab is additionally governed by a Radiation Work Permit, answer the questions in the “Radiation Work Permits” section.
      13. Complete other other inspections as necessary.
    3. Conducting Contamination Survey
      1. Mark the spots on the survey map that you will smear.
      2. Smear at least 100 cm2 per smear
      3. Hit
        1. doors/door handles
        2. refrigerators (inside refrigerator, freezer, and exterior of door(s))
        3. lab work areas
        4. floors in front of work benches
        5. cabinets where waste is stored, and
        6. RAM hoods.
          1. For active labs the smear order is:
            1. Bottom
            2. Left
            3. Right
            4. Back
            5. Sash interior
            6. Bottom Lip (Apron) and Sash Handle
            7. Exterior Glass and Control Surfaces
          2. For inactive labs, generally one smear outside, one smear inside
    4. Survey with portable meter
      1. This radiation level survey is conducted in labs where the emitted radiation is energetic enough to potentially cause exposures above background at locations where Radiation Workers can stand in the lab. Survey locations where a portion of a person’s whole body (as defined for dose and dosimetry) can be near storage refrigerators/freezers/cabinets/waste containers/etc.  If a reading ≥ 3x background is measured, document the reading on the survey map and check “No”.  Otherwise, check “Yes” at the “External radiation levels of all areas < 3x bkg?” portion of the map
        1. The appropriate meter is based on the isotopes that the PI is working with and is on the map.  Document the meter used in the applicable fields at the top of RS-117, RS-117G, or RS-118, as applicable.
        2. General logic:
          1. H-3 and C-14 radiation levels aren’t measured due to low energy of the emitted beta particles
          2. P-32 and other high energy beta emitters are detectable with a GM pancake probe
          3.  alpha-sources (Th and U-235) with gamma ray and mid to high energy beta emissions are detectable with a GM.  The alphas are only detectable if unshielded and within a few centimeters of the probe.
          4. Neutron sources require a neutron detector
          5. Ba-133 and other isotopes for which gamma ray emissions are the primary concern require an Ion Chamber if GM readings are above 3x bkg

Analyzing Smears and Recording/Reporting Smear and Survey Results

  1. Analyzing the Contamination Survey

    1. Upon return from a campus survey the smears must be counted on the appropriate machine.
    2. Must be done the same day the samples were collected so that identified contamination issues can be promptly addressed
    3. Preparing Smears for Counting
      1. LSC Smears
        1. Place 20 mL LSC vials in rack(s)
        2. Place smears in vials and add 10 mL fluor
        3. Label the first vial cap with the Building/Room in which the smears were taken, the date, and the number “1”.
        4. Label the subsequent vials sequentially with “2”, “3”, etc.
        5. If there are more than 12 vials to run for a given protocol, label the first vial of each additional rack as indicated for the very first vial in the first rack, but with the number “13” for the second rack, “25” for the third, etc.
      2. LB5100 Smears
        1. Place the smears with the “dirty”, smeared side up in the planchet in the trays.  When stacking the trays for loading into the LB 5100, stack from the bottom up starting with the first smear.  Be sure to use sequentially numbered trays when possible.
      3. Gamma Counter Smears
        1. Prep the Smears
          1. Insert the gamma counter vials (located below the gamma counter) into the appropriate rack, loading the rack slots from right to left.  This is opposite how the racks are loaded for the LSCs. Appropriate means that the Rack Number corresponds to the protocol number for the RAM of concern. Protocol numbers are shown in the photo 
            Cobra Protocol Summary on Monitor
          2. Roll the smear and insert in the vial. Do not put in scintillation fluid.
          3. Run background protocol, protocol one, first, it is only necessary to run one once a day. If one is running multiple surveys, make photocopies of the background report
          4. Use the flag that is marked on gamma counter computer
        2. After the gamma counter is run, print the report:
          1. On the printer, press "FF/Load" once
          2. Press "Tear" once. You will see the perforation just above the guide.Image of dot matrix printer and paper
          3. Tear off the paper
          4. Press "Tear" once
          5. The paper is now reset for the next protocol runImage of reset dot matrix printer
  1. Reporting Results of the Contamination Survey

    1. LSC Reports
      1. LSC survey results are recorded in an Excel Workbook “ORS Files\Campus Survey\ RS-118 (LSC Rooms)” that was used to print the maps originally.
      2. Find the appropriate map corresponding to the measured survey report
      3. Update the map header information:
        1. Date information should correspond to the date that the survey was conducted; again all efforts should be made to run the smears the day they were collected
        2. For the LSC’s, make sure to update the LSC  Machine name and Serial number, because these can and are run on different machines from quarter to quarter
        3. For the Channel Background
          1. Find the LSC Daily entry that corresponds to the day that the contamination survey report was counted on the LSC.
          2. Refer to the “Region Counted”. It will either say, “A”, “A+B”, or “A+B+C”
            1. Region A = 0-15.6 keV, Region B=15.6 keV to 256 keV, Region C-256 keV to 2000 keV.
            2. If the “Region Counted” entry is channel A, which means the radiation source is low energy (like H-3), then enter the background for just channel A, and the Channel A Efficiency (ie the tritium efficiency). This is used because it is conservative
            3. If the “Region Counted” entry is A+B, or A+B+C, then enter the daily background values from the respective channels in equation form and enter the C-14 efficiency for the channel efficiency.
              1. Example. For 8/3/2020 LSC A’s daily report was:LSC Daily Summary
                1. If the “Region Counted” was A, then Channel Background would be 8.6 and Channel Efficiency would be 55.47%
                2. If the “Region Counted” was A+B, then Channel Background would be “=8.6+18.6” which will result in 27.2. The Channel Efficiency would be 95.91%
                3. If the “Region Counted” was A+B+C, then Channel Background would be “=8.6+18.6+11.5” which will result in 38.7. The Channel Efficiency would again be 95.91%
      4. Enter Portable Survey Meter information if survey was required
        1. Type of meter
        2. Serial number
        3. Calibration due
        4. Background
        5. Units
      5. Update the digital map in the Excel file to reflect the handwritten map by moving the location bubbles appropriately
      6. Enter the report results in the digital Survey map. Enter the results data based on the “Region Counted”, A, A+B, or A+B+C. Make sure to enter it in equation form, i.e. “=A_value+B_value+C_value”. This makes it easier to check for inaccurate entries.
    2. LB5100 Reports
      LB5100 report
      1. Sign the LB5100 Sample Report
      2. Record LB5100 smear results in “ORS Files\Campus Survey\RS-117 (LB5100 Rooms)”
        1. Update the header information
          Survey report header

          1. Date information should correspond to the date that the survey was conducted; again, all efforts should be made to run the smears the day they were collected
          2. Confirm the LB5100 calibration date and update if necessary
          3. Enter alpha and beta background and efficiency information from the LB5100 header information
          4. Enter Portable Survey Meter information if survey was required
            1. Type of meter
            2. Serial number
            3. Calibration due
            4. Background
            5. Units
        2. Update the survey bubble location on the map to reflect where the actual surveys were taken
        3. Modify the data table shown below as needed to record the data
        4. Enter the starting tray number shown in the smear report above in the tray location.
          1. Usually, samples were run in sequential trays so only one tray number entry (shown in red) is necessary to update.
          2. Sometimes the actual trays are not sequential and so the tray counts in the table will have to be hand edited.
          3. If the tray number were not sequential previously, and are this time, reset the cells to automatically calculate the number based on the previous tray count.
            LB5100 Results Summary
             
          4. Enter LB5100 alpha and beta results as shown above.
    3. Gamma Counter Reports
  2. Generate Inspection Letters

    1. In EHSA database, click on the “Inspections” button
    2. In the “Inspections” box, select RAM Inspections
    3. Select the “RAM Campus Survey Results Letter - Active Labs”
    4. Select the first date of the month of the campus surveys
    5. Select the last date of the month of the campus surveys
    6. Run the report
    7. Print the report
    8. Attach to Work Order Package
  3. Final Survey Map/Results Summary

    1. Print the completed survey map/results summary for each lab.
  4. Complete the Work Order

    1. Print cover sheet for Work Order 074.
    2. Collate the sheets for the Work Order as follows
      1. For each survey type (LSC, LB5100, and/or Gamma Counter) attach:
        1. Final Survey Map
        2. Contamination results
      2. Inspection Letters
      3. The map used for the original survey
    3. Turn in completed work order to an ORS staff member for review; after review is complete, the work order will be forwarded to the RSO for final review.