1. Click “Inventory” on top row
  2. Select “Sealed Source Inventory”
  3. Select appropriate sealed source
  4. Click “Sealed Source” in the "Sealed Source Information" column.
  5. In the "Sealed Source History" section of the screen, click “+Add”
  6. Enter leak test information:
    1. Date leak test was conducted (not counted)
    2. Status = "Complete"
    3. History Type = "Leak Test" for alpha, "Radon Emanation Test" for Ra-226, or "Leak Test - 6" for beta/gamma
    4. Frequency = "6" months for beta/gamma, "3" months for alpha and Ra-226 (this should autopopulate based on the history type selected.
    5. Activity in uCi
      • ​An Excel sheet ("Leak Test Activity Calculator") is located in ORS Files > How To
    6. Next due should auto populate
    7. Result = "Pass"
      • For regular leak tests, passing means < 0.005 uCi
      • For radon emanation test, passing means < 0.001 uCi
        • Type “Radon emanation leak test; <0.001 uCi” in the Comments field. DON’T FORGET TO ALSO ENTER THE REGULAR LEAK TEST INFORMATION FOR THE RA-226 AS A SEPARATE HISTORY ITEM.
      • IF IT DID NOT PASS, INFORM RSO IMMEDIATELY
    8. Performed By (person who actually smeared it, written on the envelope)
    9. Equipment Used - Select the equipment used to analyze the leak test from the "Survey Instrument" drop-down list.
  7. Click "Save"