Click “Sealed Source” in the "Sealed Source Information" column.
In the "Sealed Source History" section of the screen, click “+Add”
Enter leak test information:
Date leak test was conducted (not counted)
Status = "Complete"
History Type = "Leak Test" for alpha, "Radon Emanation Test" for Ra-226, or "Leak Test - 6" for beta/gamma
Frequency = "6" months for beta/gamma, "3" months for alpha and Ra-226 (this should autopopulate based on the history type selected.
Activity in uCi
An Excel sheet ("Leak Test Activity Calculator") is located in ORS Files > How To
Next due should auto populate
Result = "Pass"
For regular leak tests, passing means < 0.005 uCi
For radon emanation test, passing means < 0.001 uCi
Type “Radon emanation leak test; <0.001 uCi” in the Comments field. DON’T FORGET TO ALSO ENTER THE REGULAR LEAK TEST INFORMATION FOR THE RA-226 AS A SEPARATE HISTORY ITEM.
IF IT DID NOT PASS, INFORM RSO IMMEDIATELY
Performed By (person who actually smeared it, written on the envelope)
Equipment Used - Select the equipment used to analyze the leak test from the "Survey Instrument" drop-down list.